SPECIFICATIONS
Technical Specification | Description |
Test method | ASTM D3241, IP 323, ISO 6249 Up to 21 programmable test methods |
Test Temperature | Programmable from 100°C up to 380°C |
Differential Pressure | 0 – 750 mmHg (automatically bypassed at +700 mmHg) |
Test time Range | Programmable from 4 to 600 minutes |
Fuel Aeration | Programmable time, automatic air flow control |
Aeration Flow rate | Programmable or according to test method, 1.5 L / min |
Air Flow Humidity | Humidity sensor with displayed message for dessicant replacement |
Heater tube temperature measurement | Thermocouple Type K, class 1 |
Fuel Flow | Dual 5 mL syringe, maintenance free, programmable fuel flow rate or according to test method Accuracy ±1% No flow pulse or peak |
Bus Bar Temperature Control | Independent and programmable, each bus bar temperature can be adjusted No liquid cooling circulation Peltier modules + heat pipe technology |
Heater tube section assembling | Special gauge to quickly and perfectly position the heater tube |
Fuel vapor handling | Dedicated compartment with sliding doors acting like fume hood encompassing beakers and heater tube section, connectable to a fume extractor |
Diagnostic and Service | Dedicated service menu with a flow chart layout with ability to click on each symbolic element to check its operation |
DR10 – ITR connectivity | Via Ethernet port RJ45 Full test results can be automatically transferred from the DR10 to the TO10 |
Results storage | Result database Limited only to capacity of external device |
LAN connectivity | Ethernet port RJ45 |
Printer output | USB (printer is optional) |
Data output | USB (2), Ethernet |
Dimensions (mm) | 440 x 600 x 670 (17″ x 23″ x 26″) |
Weight | 60 kg (133lbs.) |
Electrical | 100 to 240V – 8 A – 50/60Hz |
Operating temperature | From +10 to +35°C |
Relative humidity | 20% to 90% non-condensing |